USB 2.0 DAQ measurement system with simultaneous acquisition of the analog digital and counter/counter.
The resolution of the analog inputs is 16 bits. The sampling rate per channel is 3MHz, the input voltage range is ±10 V, the sum sampling rate is 6 MHz for 2 channels. The inputs are ground referenced together.
Overvoltage input protection of analog inputs: ±40 V when on or off.
4-channel D/A converter, conversion rate 10 kHz 16 bit, output voltage range ±10 V.
Synchronous acquisition of digital inputs with analog values and counters.
External trigger & clock. 2 event counters/pulse widths/pulse width measurement/incremental counters.
2 x 32 bit universal counters see technical data, pulse, frequency (resolution 10Hz), pulse width, period width/incremental counter resolution 20ns or single channel 32 bit incremental measurement, 1x, 2x and 4x interpolation programmable, switchable zero detection, timestamp resolution 100ns, the timestamp allow extremely accurate speed measurement.
Analog inputs/outputs are galvanically isolated from the controller, isolation voltage 500 volts.
Digital port: 8 opto-decoupled inputs (Ue. 2,4 -30 Volt) and 8 opto-decoupled outputs, for direct control from relay up to 60mA per output, isolation voltage 1000 Volt.
The supply voltage of the digital outputs is supplied externally in the range of 3.3 to 30 volts, isolation voltage 1000 volts.
MultiChoice in a sturdy, attractive cast aluminum housing.
Analog signal connections via BNC sockets on the housing, digital signals via removable Weidmüller screw terminals.
Including power supply unit for voltage supply 220V AC / 5V DC.
Technical data see below.